HI83399 Water & Wastewater Multiparameter (with COD) Photometer and pH meter


Water & Wastewater Multiparameter (with COD) Photometer and pH meter 

 

HI83399 is a compact, multiparameter photometer for measuring key water and wastewater quality parameters. The meter is one of the most advanced photometers available with a innovative optical design that utilizes a reference detector and focusing lens to eliminate errors from changes in the light source and from imperfections in the glass cuvette. 40 key water and wastewater quality parameters with 73 different methods covering multiple ranges are programmed into the meter. Wastewater treatment digestion parameters include COD, Total Nitrogen and Total Phosphorous which are important for monitoring nutrient removal. The HI83399 also offers an absorbance measurement mode for performance verification and for users that would like to develop their own concentration versus absorbance curves.

 

pH Specifications

pH Range Photometer: 6.5 to 8.5 pH 
pH electrode: -2.00 to 16.00 pH
pH Resolution Photometer: 0.1 pH 
pH electrode: 0.01 pH
pH Accuracy Photometer: ±0.1 pH 
pH electrode: ±0.01 pH
pH Calibration Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01)
pH Temperature Compensation Automatic (-5.0 to 100.0 oC; 23.0 to 212.0 oF); limits reduced based on the pH electrode used
pH CAL Check (electrode diagnostics) clean electrode and check buffer/check probe displayed during calibration
pH Method Photometer: phenol red
pH-mV Range ±1000 mV
pH-mV Resolution 0.1 mV
pH-mV Accuracy ±0.2 mV

Oxygen, Dissolved Specifications

Oxygen, Dissolved Range 0.0 to 10.0 mg/L (as O2)
Oxygen, Dissolved Resolution 0.1 mg/L
Oxygen, Dissolved Accuracy ±0.4 mg/L ±3% of reading
Oxygen, Dissolved Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified Winkler method

Absorbance Specifications

Absorbance Range 0.000 to 4.000 Abs
Absorbance Resolution 0.001 Abs
Absorbance Accuracy +/-0.003Abs @ 1.000 Abs

Alkalinity

Alkalinity Range Freshwater: 0 to 500 mg/L (as CaCO3
Seawater: 0 to 300 mg/L (as CaCO3)
Alkalinity Resolution 1 mg/L
Alkalinity Accuracy ±5 mg/L ±5% of reading
Alkalinity Method Colorimetric method

Aluminum Specifications

Aluminum Range 0.00 to 1.00 mg/L (as Al3+)
Aluminum Resolution 0.01 mg/L
Aluminum Accuracy ±0.04 mg/L ±4% of reading
Aluminum Method Adaptation of the aluminon method

Ammonia Specifications

Ammonia Range Low Range: 0.00 to 3.00 mg/L
Low Range (16 mm vial): 0.00 to 3.00 mg/L
Medium Range: 0.00 to 10.00 mg/L 
High Range: 0.0 to 100.0 mg/L 
High Range (16 mm vial): 0.0 to 100.0 mg/L 
(all as NH3-N)
Ammonia Resolution Low and Medium Range: 0.01 mg/L 
High Range: 0.1 mg/L
Ammonia Accuracy Low Range: ±0.04 mg/L ±4% of reading 
Low Range (16 mm vial): ±0.10 mg/L or ±5% of reading 
Medium Range: ±0.05 mg/L ±5% of reading 
High range: ±0.5 mg/L ±5% of reading
High range (16 mm vial): ±1 mg/L or ±5% of reading
Ammonia Method Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method

Anionic Surfactants Specifications

Anionic Surfactants Range 0.00 to 3.50 mg/L (as SDBS)
Anionic Surfactants Resolution 0.01 mg/L
Anionic Surfactants Accuracy ±0.04 mg/L ±3% of reading
Anionic Surfactants Method Adaptation of the USEPA method 425.1 and Standard Methods for the Examination of Water and Wastewater, 20th edition, 5540C, Anionic Surfactants as MBAS

Bromine Specifications

Bromine Range 0.00 to 8.00 mg/L (as Br2)
Bromine Resolution 0.01 mg/L
Bromine Accuracy ±0.08 mg/L ±3% of reading
Bromine Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method.

Calcium Specifications

Calcium Range Freshwater: 0 to 400 mg/L (as Ca2+)
Seawater: 200 to 600 mg/L (as Ca2+)
Calcium Resolution 1 mg/L
Calcium Accuracy Freshwater: ±10 mg/L ±5% of reading
Seawater: ±6% of reading
Calcium Method Freshwater: Adaptation of the Oxalate method 
Seawater: Adaptation of the Zincon method

Chemical Oxygen Demand Specifications

Chemical Oxygen Demand Range Low Range: 0 to 150 mg/L
Medium Range: 0 to 1500 mg/L 
High Range:0 to 15000 mg/L
Chemical Oxygen Demand Resolution 1 mg/L
Chemical Oxygen Demand Accuracy Low Range: ±5 mg/L ±4% of reading 
Medium Range: ±15 mg/L ±4% of reading 
High Range: ±150 mg/L ±2% of reading
Chemical Oxygen Demand Method Adaptation of the USEPA 410.4
ISO dichromate methods 
Mercury-free dichromate green method (LR & MR); dichromate method (HR)

Chlorine Dioxide Specifications

Chlorine Dioxide Range 0.00 to 2.00 mg/L (as ClO2)
Chlorine Dioxide Resolution 0.01 mg/L
Chlorine Dioxide Accuracy ±0.10 mg/L ±5% of reading
Chlorine Dioxide Method Adaptation of the Chlorophenol Red method.

Chlorine Specifications

Free Chlorine Range 0.00 to 5.00 mg/L (as Cl2)
Ultra Low Range: 0.000 to 0.500 mg/L (as Cl2)
Free Chlorine Resolution 0.01 mg/L 
Ultra Low Range: 0.001 mg/L
Free Chlorine Accuracy ±0.03 mg/L ±3% of reading 
Ultra Low Range: ±0.020 mg/L ±3% of reading
Total Chlorine Range 0.00 to 5.00 mg/L (as Cl2)
Ultra Low Range: 0.000 to 0.500 mg/L (as Cl2
Ultra High Range: 0 to 500 mg/L (as Cl2)
Total Chlorine Resolution 0.01 mg/L 
Ultra Low Range: 0.001 mg/L 
Ultra High Range:1 mg/L
Total Chlorine Accuracy ±0.03 mg/L ±3% of reading
Ultra Low Range: ±0.020 mg/L ±3% of reading 
Ultra High Range: ±3 mg/L ±3% of reading
Chlorine Method Adaptation of the EPA 330.5 DPD method 
Free Chlorine (ULR) & Total Chlorine (UHR): Adaptation of the Standard Methods for Examination of Water and Wastewater, 20th edition, 4500-Cl

Chromium, Hexavalent Specifications

Chromium, Hexavalent Range Low Range: 0 to 300 μg/L (as Cr6+
High Range: 0 to 1000 μg/L (as Cr6+)
Chromium, Hexavalent Resolution 1 μg/L
Chromium, Hexavalent Accuracy Low Range: ±1 μg/L ±4% of reading 
High Range: ±5 μg/L ±4% of reading
Chromium, Hexavalent Method Adaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method.

Color, Water Specifications

Color, Water Range 0 to 500 PCU (Platinum Cobalt Units)
Color, Water Resolution 1 PCU
Color, Water Accuracy ±10 PCU ±5% of reading
Color, Water Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Colorimetric Platinum Cobalt method.

Copper Specifications

Copper Range Low Range: 0.000 to 1.500 mg/L (as Cu2+)
High range: 0.00 to 5.00 mg/L (as Cu2+)
Copper Resolution 0.001 mg/L; 0.01 mg/L
Copper Accuracy Low Range: ±0.01 mg/L ±5% of reading 
High Range ±0.02 mg/L ±4% of reading
Copper Method Adaptation of the EPA bicinchoninate method

Cyanuric Acid Specifications

Cyanuric Acid Range 0 to 80 mg/L (as CYA)
Cyanuric Acid Resolution 1 mg/L
Cyanuric Acid Accuracy ±1 mg/L ±15% of reading
Cyanuric Acid Method Adaptation of the turbidimetric method

Fluoride Specifications

Fluoride Range Low Range: 0.00 to 2.00 mg/L (as F-
High range: 0.0 to 20.0 mg/L (as F-)
Fluoride Resolution 0.01 mg/L; 0.1 mg/L
Fluoride Accuracy Low Range: ±0.03 mg/L ±3% of reading 
High Range: ±0.5 mg/L ±3% of reading
Fluoride Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, SPADNS method

Hardness, Total Specifications

Hardness, Total Range Low Range: 0 to 250 mg/L (as CaCO3)
Medium Range: 200 to 500 mg/L (as CaCO3)
High Range: 400 to 750 mg/L (as CaCO3)
Hardness, Total Resolution 1 mg/L
Hardness, Total Accuracy Low Range: ±5 mg/L ±4% of reading 
Medium Range: ±7 mg/L ±3% of reading 
High Range: ±10 mg/L ±2% of reading
Hardness, Total Method Adaptation of the EPA recommended method 130.1

Hardness, Calcium Specifications

Hardness, Calcium Range 0.00 to 2.70 mg/L (as CaCO3)
Hardness, Calcium Resolution 0.01 mg/L
Hardness, Calcium Accuracy ±0.11 mg/L ±5% of reading
Hardness, Calcium Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Calmagite method

Hardness, Magnesium Specifications

Hardness, Magnesium Range 0.00 to 2.00 mg/L (as CaCO3)
Hardness, Magnesium Resolution 0.01 mg/L
Hardness, Magnesium Accuracy ±0.11 mg/L ±5% of reading
Hardness, Magnesium Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, EDTA Colorimetric method

Hydrazine Specifications

Hydrazine Range 0 to 400 μg/L (as N2H4)
Hydrazine Resolution 1 μg/L
Hydrazine Accuracy ±4% of full scale reading
Hydrazine Method Adaptation of the ASTM Manual of Water and Environmental Technology, method D1385-88, p-Dimethylaminobenzaldehyde method

Iodine Specifications

Iodine Range 0.0 to 12.5 mg/L (as I2)
Iodine Resolution 0.1 mg/L
Iodine Accuracy ±0.1 mg/L ±5% of reading
Iodine Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method

Iron Specifications

Iron Range Low Range: 0.000 to 1.600 mg/L (as Fe)
High Range: 0.00 to 5.00 mg/L (as Fe)
Iron Resolution 0.001 mg/L; 0.01 mg/L
Iron Accuracy Low Range: ±0.01 mg/L ±8% of reading 
High Range: ±0.04 mg/L ±2% of reading
Iron Method Low Range: Adaptation of the TPTZ Method 
High Range: Adaptation of the EPA Phenanthroline method 315B, for natural and treated waters

Magnesium Specifications

Magnesium Range 0 to 150 mg/L (as Mg2+)
Magnesium Resolution 1 mg/L
Magnesium Accuracy ±5 mg/L ±3% of reading
Magnesium Method Adaptation of the Calmagite method

Manganese Specifications

Manganese Range Low Range: 0 to 300 μg/L (as Mn)
High Range: 0.0 to 20.0 (as Mn)
Manganese Resolution 1 μg/L; 0.1 mg/L
Manganese Accuracy Low Range: ±10 μg/L ±3% of reading 
High Range: ±0.2 mg/L ±3% of reading
Manganese Method Low Range: Adaptation of the PAN Method
High Range: Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Periodate method

Molybdenum Specifications

Molybdenum Range 0.0 to 40.0 mg/L (as Mo6+)
Molybdenum Resolution 0.1 mg/L
Molybdenum Accuracy ±0.3 mg/L ±5% of reading
Molybdenum Method Adaptation of the mercaptoacetic acid method

Nickel Specifications

Nickel Range Low Range: 0.000 to 1.000 mg/L (as Ni) 
High Range: 0.00 to 7.00 g/L (as Ni)
Nickel Resolution 0.001 mg/L; 0.01 g/L
Nickel Accuracy Low range: ±0.010 mg/L ±7% of reading 
High Range: ±0.07 g/L ±4% of reading
Nickel Method Low Range: Adaptation of the PAN method 
High Range: Adaptation of the photometric method

Nitrate Specifications

Nitrate Range 0.0 to 30.0 mg/L (as NO3-- N)
16 mm vial: 0.0 to 30.0 mg/L (as NO3-- N)
Nitrate Resolution 0.1 mg/L
Nitrate Accuracy ±0.5 mg/L ±10% of reading
16 mm vial: ±1.0 mg/L or ±3% of reading
Nitrate Method Adaptation of the cadmium reduction method 
16 mm vial: Chromotropic acid method

Nitrite Specifications

Nitrite Range Freshwater 
Low Range: 0 to 600 μg/L (as NO2--N)
High Range: 0 to 150 mg/L (as NO2-)
Seawater Ultra Low Range: 0 to 200 μg/L (as NO2--N)
Nitrite Resolution Freshwater: 1 μg/L; 1 mg/L
Seawater: 1 μg/L
Nitrite Accuracy Freshwater
Low Range: ±20 μg/L ±4% of reading 
High Range: ±4 mg/L ±4% of reading 
Seawater
±10 μg/L ±4% of reading
Nitrite Method Low Range and Seawater: Adaptation of the EPA Diazotization method 354.1 
High Range: Adaptation of the Ferrous Sulfate method

Nitrogen, Total Specifications

Nitrogen, Total Range Low Range: 0.0 to 25.0 mg/L (as NO3--N)
High Range: 0 to 150 mg/L (as N)
Nitrogen, Total Resolution 0.1 mg/L; 1 mg/L
Nitrogen, Total Accuracy Low Range: ±1.0 mg/L or ±5% of reading
High Range: ±3 mg/L or ±4% of reading
Nitrogen, Total Method Chromotropic acid method

Oxygen, Scavenger Specifications

Oxygen, Scavenger Range 0 to 1000 μg/L (as DEHA) 
0.00 to 1.50 mg/L (as Carbohydrazide)
0.00 to 2.50 mg/L (as Hydroquinone)
0.00 to 4.50 mg/L (as ISO-ascorbic acid)
Oxygen, Scavenger Resolution 1 μg/L (DEHA); 0.01 mg/L
Oxygen, Scavenger Accuracy ±5 μg/L ±5% of reading
Oxygen, Scavenger Method Adaptation of the iron reduction method

Ozone Specifications

Ozone Range 0.00 to 2.00 mg/L (as O3)
Ozone Resolution 0.01 mg/L
Ozone Accuracy ±0.02 mg/L ±3% of reading
Ozone Method Colorimetric DPD Method

Phosphate Specifications

Phosphate Range Freshwater
Low Range: 0.00 to 2.50 mg/L (as PO43-)
High range: 0.0 to 30.0 mg/L(as PO43-)
Seawater Ultra Low Range: 0 to 200 μg/L (as P)
Phosphate Resolution Freshwater: 0.01 mg/L; 0.1 mg/L
Seawater: 1 μg/L
Phosphate Accuracy Freshwater
Low Range: ±0.04 mg/L ±4% of reading 
High Range: ±1 mg/L ±4% of reading 
Seawater Ultra Low Range: ±5 μg/L ±5% of reading
Phosphate Method Freshwater Low Range: Adaptation of the Ascorbic Acid method
Freshwater High Range and Seawater Ultra Low Range: Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method

Phosphorous, Acid Hydrolyzable Specifications

Phosphorous, Acid Hydrolyzable Range 0.00 to 1.60 mg/L (as P)
Phosphorous, Acid Hydrolyzable Resolution 0.01 mg/L
Phosphorous, Acid Hydrolyzable Accuracy ±0.05 mg/L or ±5% of reading
Phosphorous, Acid Hydrolyzable Method Adaptation of the EPA method 365.2 and Standard Methods for the Examination of Water and Wastewater, 20th edition, 4500-P E, ascorbic acid method

Phosphorous, Reactive Specifications

Phosphorous, Reactive Range Low Range: 0.00 to 1.60 mg/L (as P) 
High Range: 0.0 to 32.6 mg/L (as P)
Phosphorous, Reactive Resolution 0.01 mg/L; 0.1 mg/L
Phosphorous, Reactive Accuracy Low Range: ±0.05 mg/L or ±4% of reading 
High Range: ±0.5 mg/L or ±4% of reading
Phosphorous, Reactive Method Low Range: Adaptation of Standard Methods for the Examination of Water and Wastewater, 20th edition, ascorbic acid method, and EPA method 365.2 
High Range: Adaptation of the vanadomolybdophosphoric acid method.

Phosphorous, Total Specifications

Phosphorous, Total Range Low Range: 0.00 to 1.15 mg/L (as P) 
High Range: 0.0 to 32.6 mg/L (as P)
Phosphorous, Total Accuracy Low Range: ±0.05 mg/L or ±6% of reading 
High Range: ±0.5 mg/L or ±5% of reading
Phosphorous, Total Resolution Low Range: 0.01 mg/L; 0.1 mg/L
Phosphorous, Total Method Low Range: Adaptation of Standard Methods for the Examination of Water and Wastewater, 20th edition, 4500-P E, ascorbic acid method, and EPA method 365.2 
High Range: Adaptation of the vanadomolybdophosphoric acid method

Potassium Specifications

Potassium Range 0.0 to 20.0 mg/L (as K)
Potassium Resolution 0.1 mg/L
Potassium Accuracy ±3.0 mg/L ±7% of reading
Potassium Method Adaptation of the Turbidimetric Tetraphenylborate method

Silica Specifications

Silica Range Low Range: 0.00 to 2.00 mg/L (as SiO2)
High Range: 0 to 200 mg/L (as SiO2)
Silica Resolution 0.01 mg/L; 1 mg/L
Silica Accuracy Low Range: ±0.03 mg/L ±3% of reading
High Range: ±1 mg/L ±5% of reading
Silica Method Low Range: Adaptation of the ASTM Manual of Water and Environmental Technology, D859, Heteropoly Molybdenum Blue method 
High Range: Adaptation of the USEPA Method 370.1 and Standard Method 4500-SiO2

Silver Specifications

Silver Range 0.000 to 1.000 mg/L (as Ag)
Silver Resolution 0.001 mg/L
Silver Accuracy ±0.020 mg/L ±5% of reading
Silver Method Adaptation of the PAN method

Sulfate Specifications

Sulfate Range 0 to 150 mg/L (as SO42-)
Sulfate Resolution 1 mg/L
Sulfate Accuracy ±5 mg/L ±3% of reading
Sulfate Method Turbidimetric - Sulfate is precipitated with barium chloride crystals

Zinc Specifications

Zinc Range 0.00 to 3.00 mg/L (as Zn)
Zinc Resolution 0.01 mg/L
Zinc Accuracy ±0.03 mg/L ±3% of reading
Zinc Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Zincon method

General Meter Specifications

Input Channels 1 pH electrode input and 5 photometer wavelengths
pH Electrode digital pH electrode (not included)
Logging Type log on demand with user name and sample ID optional input
Logging Memory 1000 readings
Connectivity USB-A host for flash drive; micro-USB-B for power and computer connectivity
GLP calibration data for connected pH electrode
Display 128 x 64 pixel LCD with backlight
Battery Type / Life 3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement
Power Supply 5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included)
Environment 0 to 50.0 oC (32 to 122.0 oF); 0 to 95% RH, non-condensing
Dimensions 206 x 177 x 97 mm (8.1 x 7.0 x 3.8")
Weight 1.0 kg (2.2 lbs.)
Photometer/Colorimeter Light Source 5 LEDs with 420 nm, 466 nm, 525 nm, 575 nm, and 610 nm narrow band interference filters
Photometer/Colorimeter Light Detector silicon photodetector
Bandpass Filter Bandwidth 8 nm
Bandpass Filter Wavelength Accuracy ±1 nm
Cuvette Type round, 24.6 mm
Number of Methods 128 max.
Ordering Information HI83399 is supplied with sample cuvettes and caps (4 ea.), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual.
Visitors: 137,686