เครื่องวัดความขุ่น HANNA HI98713

 



The HI98713 Precision ISO Turbidity Portable Meter is specially designed for water quality measurements, providing reliable and accurate readings, even within low turbidity ranges. The instrument is based on a state-of-the-art optical system which guarantees accurate results, assures long term stability, and minimizes stray light and color interferences. Periodic calibration with the supplied standards compensates for any variations in intensity of the LED light source.  The 25 mm round cuvettes composed of special optical glass guarantee the repeatability of turbidity measurements


Features at-a-glance

Ratio Measurement Mode - The HI98713 measures turbidity using the ratio method with a 90° and 180° light detector for more accurate measurements. Turbidity measurements can be made in the 0.00 to 1000 FNU range.

Multiple reading modes
 - Normal measurement, continuous measurement, or signal averaging measurement are reading modes available.

ISO Compliant -  The HI98713 meets and exceeds the requirements of ISO 7027 method for turbidity measurements by use of an infrared LED light source.

Calibration – The HI98713 has a powerful calibration function that compensates for variation in light intensity. The calibration can be done using the supplied calibration solutions or user-prepared standards. A two, three, or four-point turbidity calibration can be performed by using the supplied (<0.1, 15, 100 and 750 FNU) standards. Calibration points can be modified if user-prepared standards are used.

AMCO AEPA-1 Primary Turbidity Standard - The AMCO AEPA-1 supplied standards are recognized as a primary standard by the USEPA. These non-toxic standards are made of styrene divinylbenzene polymer spheres that are uniform in size and density. The standards are reusable and stable with a long shelf life.

Fast Tracker - For advanced ˜field applications, the HI98713 is equipped with Fast Tracker - Tag Identi˜fication System (T.I.S.) that makes data collecting and management simpler than ever. Fast Tracker allows users to record the time and location of a speci˜fic measurement or series of measurements using iButton® tags near sampling points for quick and easy readings. Each iButton® tag contains a computer chip with a unique identi˜cation code encased in stainless steel.

GLP Data – The HI98713 features complete GLP (Good Laboratory Practice) functions that allow traceability of the calibration conditions. Data includes calibration points, date, and time.

Data Logging – Up to 200 measurements can be stored in the internal memory and recalled at any time.

Data Transfer – For further storage or analysis options, logged data can be downloaded to a Windows® compatible PC using the USB or RS232 port and the HI92000 software. 

Backlit Graphic LCD Display – A graphic LCD display provides an easy to understand, user-friendly interface. All messages are in plain text making them easy to read.


Turbidity Specifications

Range 0.00 to 9.99; 10.0 to 99.9; 100 to 1000 NTU
Resolution 0.01; 0.1; 1 FNU
Range Selection automatic
Accuracy ±2% of reading plus 0.1 FNU
Repeatability ±1% of reading or 0.01 FNU, whichever is greater
Stray Light < 0.1 FNU
Light Detector silicon photocell
Method ratio nephelometric method (90° & 180°)
Measuring Mode normal, average, continuous
Turbidity Standards < 0.1, 15, 100 and 750 FNU
Calibration two, three, or four-point calibration
General Specification

Light Source 860 nm IR LED / greater than 100,000 readings
Log Memory 200 records
Connectivity USB or RS232
Environment 0 to 50°C (32 to 122°F), RH max 95% non-condensing
Power Supply 1.5V AA batteries (4) / 12 VDC adapter; auto-off after 15 minutes of non-use
Dimensions 224 x 87 x 77 mm (8.8 x 3.4 x 3.0”)
Weight 512 g (18 oz.)
Ordering Information HI98713 is supplied with sample cuvettes and caps (5), HI98713-11 calibration cuvettes, silicone oil, cuvette wiping cloth, batteries, AC adapter, instructions, and rugged carrying case.
Warranty 2 years

 

 

 